can some body email me a standard golden nanoparticle image please
In addition to the other advice already given:
1) Reduce the physical size of your sample as small as possible. More mass = more thermal vibrations.
2) Reduce the temperature of your sample (and room) as much as feasible and comfortably possible.
3) Dont place your AFM on a table. It should be mounted on a heavy slab which is in turn suspended by a system of springs, much like a DJ spinning board for techno music. If you dont have such a setup, use the heaviest table you can find.
4) If possible, use an adhesive to attach your sample to a holding chip, but use as little adhesive as possible. Allow at least 24 hours after affixing with the adhesive before imaging, to relieve tension.
5) If possible, load the AFM, focus, and then allow it to sit for several hours (or an entire day) in order to come to total thermal equilibrium and reduce thermal drift and vibrations.
its a good information.
i will follow these things in future..
i want to know more about cell sample observation...
First of all look at the the mode of operation. If it is contact mode then try to use cantilevers having the lowest force constant. Tapping mode works the best for all biological samples. If you have problem during scanning u can always remove cantilever and put it in ethanol for few minutes before using it again.
second thing that is important is to find a stable environment to avoid
vibrations which can cause noise to any scanning probe micrscope. You can avoid this by closing the door of your room and asking everybody not to enter the room during measurements. Another important point is the electrical noise. So AFM power connection should be connected only to one socket and other instrument must not be connected in to that socket.
Regarding during experiments try to play with your servo gain. too high or low value of servo gain can always result in noise.
Hope this helps. If u have any specific problem then let me know.. These are just generalised problem associated with scanning probe microscopes.
best wishes..
thanks for getting back to me..
i am using asylum AFM
one BIO-PD...
not their latest saphire..
regards
bgnair
What mode of AFM are you using,,, could u eleborate ? Then only I will be able to comment about noise....
Is there any researchers who are specialized in AFM or SPM on biological sample.I am experiencing noises when i go down to a level of 200nm size scan.
rgrds
Agilent Technologies is proud to announce a brand new series of AFM e-Seminars this spring! These one-hour live events will be led by some of the most prestigious scientists in their respective fields. The unique Agilent e-Seminar format includes an open Q&A session in which all online attendees are welcome to query the presenters.
Choosing the Correct Cantilever for your Application
May 13, 2009 at 3PM GMT
Speakers: Oliver Krause, PhD NanoWorld Group
Probe Manufacturing - Why Probes Look the Way They Do.
In this seminar, Dr Krause will explain the reason why probes have many shapes, the different probe types that are available and what extra refinements can be done for various applications. This seminar will also give you a guide line of how to choose the correct probe for your application
Speaker: Arvind Raman, PhD Purdue University
Veda: a Free Online Simulation Tool for the AFM Community
VEDA (Virtual environment for dynamic AFM) is a suite of free, cyber-enabled tools available on www.nanohub.org that simulate tapping mode AFM using magnetic/acoustic excitation (MAC/AAC) on organic and inorganic surfaces in liquids as well as ambient conditions. In this seminar, a brief introduction to the tools and their use in choosing cantilevers and operating conditions for optimal imaging will be presented
Register